Device-related differences in blood pressure measurement and their impact on hypertension prevalence in Korean children and adolescents

Article information

Epidemiol Health. 2026;48.e2026037
Publication date (electronic) : 2026 July 27
doi : https://doi.org/10.4178/epih.e2026037
1Department of Pediatrics, CHA Bundang Medical Center, CHA University, Seongnam, Korea
2Department of Preventive Medicine, Hanyang University College of Medicine, Seoul, Korea
3Institute for Health and Society, Hanyang University, Seoul, Korea
4Department of Internal Medicine, Boramae Medical Center, Seoul National University College of Medicine, Seoul, Korea
5Division of Cardiology, Department of Internal Medicine, Nowon Eulji Medical Center, Eulji University College of Medicine, Seoul, Korea
6Division of Cardiology, Department of Internal Medicine, Ewha Womans University Seoul Hospital, Ewha Womans University College of Medicine, Seoul, Korea
7Division of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea
8Department of Internal Medicine, Seoul National University Bundang Hospital, Seoul National University College of Medicine, Seongnam, Korea
9Division of Cardiology, Department of Internal Medicine, Hanyang University College of Medicine, Seoul, Korea
10Division of Health and Nutrition Survey and Analysis, Bureau of Chronic Disease Prevention and Control, Korea Disease Control and Prevention Agency, Cheongju, Korea
11Division of Cardiology, Department of Internal Medicine, Dongguk University Ilsan Hospital, Goyang, Korea
Correspondence: Eun Mi Lee Division of Cardiology, Department of Internal Medicine, Dongguk University Ilsan Hospital, 27 Dongguk-ro, Ilsandong-gu, Goyang 10326, Korea E-mail: mdemlee@naver.com
*Kim & Kim contributed equally to this work as joint first authors.
Received 2026 April 16; Revised 2026 July 6; Accepted 2026 July 7.

Abstract

OBJECTIVES

Accurate blood pressure (BP) measurement is essential for reliably estimating hypertension prevalence in population-based surveys. However, evidence supporting the use of oscillometric BP devices in pediatric populations remains limited. This study compared BP measurements obtained using an oscillometric device (OD) and an auscultatory device (AD) and evaluated inter-device agreement and the impact of device choice on hypertension prevalence estimates.

METHODS

BP was measured sequentially with the AD and OD in 251 children and adolescents participating in the Korea National Health and Nutrition Examination Survey. Inter-device agreement was assessed using mean differences, limits of agreement, concordance correlation coefficients (CCCs), and regression analyses. Differences in prevalence estimates for elevated blood pressure (EBP) and hypertension were also examined.

RESULTS

Mean inter-device differences were minimal for systolic blood pressure (SBP; 0.2±5.9 mmHg) but negative for diastolic blood pressure (DBP; −1.5±6.1 mmHg), with larger discrepancies in older girls. Agreement was moderate to good for SBP (CCC=0.79) and weaker for DBP (CCC=0.60). Nevertheless, more than 85% of paired measurements were within ±10 mmHg. Prevalence estimates for EBP (11.6 vs. 10.0%) and hypertension (4.8 vs. 4.4%) were comparable between devices, with relatively high negative agreement and overall agreement.

CONCLUSIONS

Despite modest underestimation of DBP in specific subgroups, oscillometric BP measurement showed acceptable agreement with the auscultatory method and yielded similar hypertension prevalence estimates. These findings suggest that ODs may be useful for pediatric BP measurement in public health surveillance systems.

GRAPHICAL ABSTRACT

Key Message

In children and adolescents, the Microlife WatchBP Office AFIB oscillometric device showed acceptable agreement with the Accoson Greenlight 300 auscultatory device for systolic blood pressure, although diastolic blood pressure agreement was more variable in some subgroups. The two devices yielded generally comparable population-level estimates of elevated blood pressure and hypertension. These findings suggest that the Microlife device may be useful for pediatric blood pressure measurement in national health surveys.

INTRODUCTION

Accurate blood pressure (BP) measurement is fundamental to the diagnosis and surveillance of hypertension across the life course. In children and adolescents, reliable BP assessment is particularly important because elevated blood pressure (EBP) in youth is associated with increased cardiovascular risk in adulthood [1-3]. In large-scale population-based surveys, such as the Korea National Health and Nutrition Examination Survey (KNHANES), BP measurement must balance accuracy, feasibility, and standardization to ensure reliable prevalence estimates and comparability over time.

KNHANES has served as a key national health surveillance system since 1998. Historically, BP was measured using mercury sphygmomanometers, which were considered reference standards. After the Minamata Convention mandated the global phase-out of mercury-based instruments, KNHANES adopted the Accoson Greenlight 300TM as a mercury-free auscultatory device (AD) [4]. Although auscultatory methods are regarded as reference standards, they are subject to observer variability and are less practical for large-scale epidemiologic surveys [5].

Oscillometric devices (ODs) offer advantages in standardization and feasibility and are increasingly used in population-based studies. Current pediatric hypertension guidelines from the American Academy of Pediatrics (AAP) and the European Society of Hypertension (ESH) endorse validated ODs for BP screening in children and adolescents [6,7]. However, prior studies have reported discrepancies between oscillometric and auscultatory measurements, particularly for diastolic blood pressure (DBP), raising concerns about potential misclassification of BP status [8-11]. Moreover, most previous studies have focused primarily on adults or on inter-device agreement itself, whereas evidence on the impact of these differences on population-level estimates of EBP and hypertension in pediatric populations remains limited.

To address these limitations, the Microlife WatchBP Office AFIB OD has been introduced for use in KNHANES. Although adult studies have shown acceptable agreement with auscultatory methods [12], evidence regarding its performance in pediatric population-based settings remains limited. Physiological characteristics of children and adolescents, including arterial compliance and hemodynamic variability, may influence oscillometric BP measurements, particularly DBP. Additional evidence is therefore needed on inter-device agreement and the potential impact of device-related differences on BP classification in pediatric populations.

Accordingly, this study aimed to evaluate agreement between oscillometric and auscultatory BP measurements and to determine the impact of device-related differences on estimates of EBP and hypertension in a nationally representative sample of children and adolescents participating in KNHANES.

MATERIALS AND METHODS

Study participants

Between April 2023 and September 2023, 251 children and adolescents aged 6–18 years who participated in KNHANES were consecutively enrolled in this study. Each participant underwent BP measurement with both the Microlife WatchBP Office AFIB (OD) and the Accoson Greenlight 300TM (AD). Measurements were conducted at four mobile examination centers (MECs) as part of KNHANES [13]. Children were excluded if they did not complete six BP measurements with each device (three per session) or if their mid-arm circumference was <14 cm or >42 cm.

Device quality control

Calibration and quality control (QC) procedures for both devices followed the standardized KNHANES QC protocol [4]. Device accuracy was assessed using the Rigel BP-SIM NIBP Simulator® (Rigel Medical, Edison, NJ, USA) before and after field use. Daily and weekly QC procedures were also conducted at each MEC. Devices were required to meet predefined accuracy criteria, including pressure differences within ±3 mmHg for most paired readings. No device malfunction was identified during the study period.

Blood pressure measurement protocol

The OD had previously undergone validation studies in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO 81060-2:2013) [14,15], whereas the AD was validated according to the 2002 European Society of Hypertension International Protocol [16].

BP measurements were performed by four trained nurses at the MECs according to KNHANES guidelines [4]. After a 5-minute seated rest with back support and the arm positioned at heart level, participants underwent BP measurement with an appropriately sized cuff selected according to mid-arm circumference. Radial pulse was recorded for 30 seconds in participants with a regular rhythm and for 60 seconds in those with an irregular rhythm.

Measurements followed a fixed sequence to minimize recall bias: three consecutive readings were obtained with the AD, followed by three readings with the OD, with 30-second intervals between readings. For the AD, the first and fifth Korotkoff sounds were recorded as systolic blood pressure (SBP) and DBP, respectively. All measurements were obtained from the same arm under consistent conditions.

Definitions of blood pressure (BP) difference, absolute BP difference, and hypertension classification

The first reading from each device was excluded from the analysis, and the mean of the second and third readings was used. Differences in SBP and DBP (D-SBP and D-DBP) were defined as OD values minus AD values. Absolute differences (A-SBP and A-DBP) were calculated as the absolute values of these differences. EBP was defined as BP at or above the 90th percentile, and hypertension was defined as BP at or above the 95th percentile for age, sex, and height according to KNHANES reference data [17,18].

Reference for interpretation of the degree of agreement between two devices

The degree of agreement between the two devices was interpreted according to the 2018 Association for the AAMI/ESH/ISO Universal Standard for validation of BP measuring devices [19]. A device was considered to be within an acceptable range when the mean difference between OD and AD did not exceed 5±8 mmHg and at least 85% of paired measurements were within ±10 mmHg.

Statistical analysis

Categorical variables were expressed as counts and percentages, whereas continuous variables were presented as means±standard deviations. Differences between OD and AD were compared, correlations were assessed using Pearson correlation coefficients, and concordance was analyzed using Lin’s concordance correlation coefficient (CCC). Bland–Altman plots were constructed to determine limits of agreement (LOA). Using AD as the comparator, the agreement and classification performance of OD were summarized using positive agreement, negative agreement, overall agreement, and prevalence-adjusted bias-adjusted kappa (PABAK). The McNemar test was used to assess whether discordant paired classifications were symmetric. Factors associated with inter-device BP differences were examined using univariable and multivariable linear regression analyses. Subgroup analyses were performed by age group, defined as 6–9 years, 10–12 years, 13–15 years, and 16–18 years.

All statistical analyses were conducted using SAS version 9.4 (SAS Institute Inc., Cary, NC, USA), and statistical significance was set at p-value <0.05.

Ethics statement

The study was approved by the Institutional Review Board of the Korea Disease Control and Prevention Agency (approval No. 2022-11-16-R-A). Written informed consent was obtained from participants and their parents.

RESULTS

Study population

A total of 251 children and adolescents were included: 121 boys (48.2%) and 130 girls (51.8%), with a mean age of 12.6±3.3 years (Table 1). Boys had greater arm circumference (24.3±4.3 vs. 22.4±3.3 cm), body weight (53.1±16.6 vs. 45.9±13.2 kg), and height (158.2±16.2 vs. 150.6±15.0 cm) than girls (all p<0.001), whereas girls had a higher pulse rate (81.4±10.2 vs. 78.0±11.7 bpm, p=0.014).

Characteristics of the study population according to sex

Inter-device blood pressure differences

SBP showed minimal inter-device differences across age groups, whereas DBP measured by the OD tended to be lower than DBP measured by the AD (Table 2, Supplementary Material 1). In the overall sample, the mean SBP difference was 0.2±5.9 mmHg (p=0.603). The mean DBP difference was −1.5±6.1 mmHg (p<0.001), with larger differences observed among adolescents aged 16–18 years.

Anthropometrics and blood pressure distributions by age

In sex-stratified analyses, SBP differences were small in both boys and girls. DBP differences were greater in girls, particularly among older adolescents (−4.0±4.4 mmHg; p<0.001).

Blood pressure (BP) agreement between BP measurement devices

Agreement between devices differed by BP component, sex, and age (Table 3). In the overall population, concordance was moderate to good for SBP (CCC, 0.79; 95% confidence interval [CI], 0.74 to 0.83) and lower for DBP (CCC, 0.60; 95% CI, 0.52 to 0.67). SBP concordance was similar in boys (CCC, 0.76) and girls (CCC, 0.77), whereas DBP concordance was 0.63 in boys and 0.56 in girls.

Correlations between AD and OD according to age and sex

Across age groups, SBP concordance ranged from 0.73 to 0.82, whereas DBP concordance remained lower; the lowest value was observed in girls aged 13–15 years (CCC, 0.43). Overall, agreement was higher for SBP than for DBP across subgroups.

Absolute difference between blood pressure measurement devices

Figure 1 shows the distribution of absolute inter-device differences. More than 85% of paired measurements for both SBP and DBP were within ±10 mmHg. In the overall population, 62.6% of SBP measurements and 57.8% of DBP measurements were within ±5 mmHg.

Figure 1.

Distribution of the absolute difference between two devices. The graphs display the percentage of participants. A-diff, absolute difference; DBP, diastolic blood pressure; SBP, systolic blood pressure.

Bland–Altman plot and limits of agreement by age and sex

Bland–Altman plots showed small mean inter-device differences in the overall population (SBP: 0.16 mmHg; DBP: −1.48 mmHg), with LOA of approximately ±12 mmHg for SBP and slightly wider ranges for DBP (Figure 2, Supplementary Material 2). Across subgroups, SBP differences remained small, whereas DBP differences were more variable, particularly in older adolescents and girls. Higher auscultatory DBP values were associated with more negative OD–AD differences.

Figure 2.

The Bland-Altman plots between OD and AD for SBP and DBP according to sex. (A) SBP in the overall population, (B) DBP in the overall population, (C) SBP in boys, (D) DBP in boys, (E) SBP in girls, and (F) DBP in girls. AD, auscultatory device (Greenlight); DBP, diastolic blood pressure; OD, oscillometric device (Microlife); SBP, systolic blood pressure.

Factors contributing to blood pressure differences

In multivariable analysis, oscillometric SBP was independently associated with SBP differences (Supplementary Material 3). For DBP, oscillometric DBP and age were independently associated with inter-device differences. The models had limited explanatory power (SBP: R2=0.174; DBP: R2=0.077), indicating that the included variables explained only a small proportion of inter-device variability.

Implications for hypertension classification

Prevalence estimates for EBP and hypertension were similar between devices (Table 4). EBP prevalence was 11.6% with the AD and 10.0% with the OD, whereas hypertension prevalence was 4.8% and 4.4%, respectively.

Prevalence of elevated blood pressure and hypertension according to age and sex

Overall agreement was 91.2% for EBP (PABAK=0.83) and 96.4% for hypertension (PABAK=0.93). Similar agreement was observed in sex-specific analyses (Table 5). In the overall sample and sex-stratified analyses, OD showed high negative agreement and overall agreement with AD for both EBP and hypertension classifications. PABAK values also indicated substantial to very high agreement. However, positive agreement was relatively modest, particularly for hypertension, and confidence intervals were wide in some subgroups, suggesting uncertainty due to the small number of positive cases. McNemar tests were not statistically significant in any comparison, indicating no evidence of systematic asymmetry in discordant classifications between OD and AD.

Diagnostic agreement between AD and OD according to sex

DISCUSSION

The principal findings of this study were as follows. First, SBP measurements showed moderate to good agreement between the OD and AD across age and sex groups. Second, DBP measured by the OD tended to be lower, particularly among older adolescent girls. Third, although inter-device differences were observed, more than 85% of paired measurements were within ±10 mmHg, and population-level estimates of EBP and hypertension were similar between devices. However, positive agreement for hypertension classification was modest, indicating that agreement at the individual diagnostic level should be interpreted cautiously.

The higher agreement observed for SBP than for DBP is consistent with previous pediatric studies and reflects known technical characteristics of oscillometric measurement [20,21]. ODs estimate SBP and DBP indirectly from oscillation amplitude envelopes, and signal detection is generally more robust for SBP than for DBP, particularly in children with smaller arterial diameters and greater hemodynamic variability [20]. In this study, SBP showed stable agreement across age and sex groups, whereas DBP differences were more variable, particularly among older adolescents.

Lower concordance for DBP, especially in adolescent girls, may reflect both physiological and methodological factors. Pubertal changes, including alterations in vascular compliance and hormonal influences, as well as differences in pulse pressure amplification, may affect oscillometric signal detection during diastole [20,21]. In addition, auscultatory measurement in children is subject to variability because of difficulty identifying Korotkoff sounds and inherent measurement variability [5,20]. These factors may contribute to greater variability in DBP measurements across devices. As AD-measured DBP increased, OD–AD differences tended to become more negative; this pattern may have contributed to the lower concordance of OD-based EBP and hypertension classification among older participants.

Despite these differences, the impact on population-level hypertension classification was limited. Prevalence estimates for EBP and hypertension were generally similar between devices, although differences in agreement were observed across age and sex subgroups. Compared with the AD, the OD showed relatively high negative agreement and high overall agreement, with no statistically significant evidence of asymmetry in discordant classifications based on the McNemar test. However, positive agreement was relatively modest, particularly for hypertension classification, and CIs were wide in some subgroups. These findings suggest that oscillometric measurements may be useful for population-level surveillance, but positive classifications should be interpreted cautiously, and the two methods should not be considered fully interchangeable for individual diagnosis. This information on diagnostic concordance between the two devices may provide a useful reference for interpreting pediatric BP classifications in nationally representative health statistics.

Current pediatric hypertension guidelines from the AAP and ESH state that EBP identified by oscillometric measurement should be confirmed by auscultatory measurement before a clinical diagnosis of hypertension is established [6,7]. In this context, the present findings are more relevant to population-based surveillance and screening than to direct diagnostic interchangeability between devices. The observed inter-device differences in DBP and modest positive agreement for hypertension classification suggest that EBP identified by oscillometric measurement should be interpreted cautiously at the individual diagnostic level.

From an epidemiologic perspective, these findings are particularly relevant to population-based surveillance. National health surveys require measurement methods that ensure both accuracy and feasibility. Although auscultatory methods remain reference standards, they require trained observers and are subject to inter-observer variability [5]. In contrast, this OD offers advantages in standardization and operational efficiency while maintaining acceptable agreement with auscultatory measurements and preserving hypertension prevalence estimates.

Several limitations should be considered. First, this study was conducted in a sample of Korean children and adolescents, which may limit generalizability to other populations. Second, variability in pediatric hypertension definitions may influence prevalence estimates. Third, the relatively low prevalence of hypertension limited detailed subgroup analyses. In addition, the similar overall prevalence estimates observed in this study may partly reflect offsetting differences across age and sex subgroups. Therefore, comparability may vary depending on the age and sex composition of the target population. Fourth, BP measurements were obtained sequentially rather than using a randomized same-arm alternating measurement design, which may have introduced order effects. In particular, if oscillometric BP measurements had been obtained first, prior awareness of the displayed values could have influenced subsequent auscultatory measurements by the observer. The 2018 AAMI/ESH/ISO Universal Standard recommends a same-arm sequential alternating measurement protocol for formal validation studies. However, because this protocol was not feasible in the present study, auscultatory BP measurement was performed first to reduce the potential influence of recall bias from automated BP readings. Finally, this study should not be interpreted as a formal validation study according to the 2018 AAMI/ESH/ISO Universal Standard because the measurement protocol did not use the same-arm sequential alternating design recommended for formal device validation. Instead, the present study was designed to evaluate inter-device agreement and the potential impact of device-related differences on BP classification and hypertension prevalence estimates within the operational framework of KNHANES.

Despite these limitations, this study has important strengths. It was conducted within a nationally representative survey using standardized protocols and rigorous QC [4], which enhances the reliability and applicability of the findings for public health surveillance.

In conclusion, oscillometric BP measurements showed acceptable agreement with auscultatory measurements for SBP, whereas DBP agreement was more variable in some pediatric subgroups. Although caution is warranted when interpreting individual hypertension classifications, population-level estimates of EBP and hypertension were generally comparable between devices. These findings suggest that ODs may be useful for pediatric BP surveillance and screening in large-scale population-based studies.

Data availability

The data generated in this study are available from the corresponding author upon reasonable request.

Supplementary materials

Supplementary materials are available at https://doi.org/10.4178/epih.e2026037.

Supplementary Material 1.

Anthropometrics and blood pressure distributions by age and sex

epih-48-e2026037-Supplementary-1.docx

Supplementary Material 2.

Limit of agreement (LOA) according to sex and age

epih-48-e2026037-Supplementary-2.docx

Supplementary Material 3.

Factors associated with inter-device BP differences

epih-48-e2026037-Supplementary-3.docx

Notes

Conflict of interest

The authors have no conflicts of interest to declare for this study.

Funding

This work was supported by the Dongguk University College of Medicine Research Fund of 2026.

Acknowledgements

None.

Author contributions

Conceptualization: Kim SH, Kim YM, Kim HL, Park SM, Cho IJ, Kim DH, Kim KI, Shin J, Oh K, Lee EM. Data curation: Kim YM, Kim SH, Lee EM. Formal analysis: Kim YM, Kim SH, Lee EM. Funding acquisition: Lee EM. Methodology: Kim YM, Kim SH, Lee EM. Project administration: Lee EM. Visualization: Kim YM, Kim SH, Lee EM. Writing – original draft: Kim YM, Kim SH, Lee EM. Writing – review & editing: Kim SH, Kim YM, Kim HL, Park SM, Cho IJ, Kim DH, Kim KI, Shin J, Oh K, Lee EM.

References

1. Vos LE, Oren A, Uiterwaal C, Gorissen WH, Grobbee DE, Bots ML. Adolescent blood pressure and blood pressure tracking into young adulthood are related to subclinical atherosclerosis: the Atherosclerosis Risk in Young Adults (ARYA) study. Am J Hypertens 2003;16:549–555. https://doi.org/10.1016/s0895-7061(03)00857-4.
2. Juhola J, Magnussen CG, Viikari JS, Kähönen M, Hutri-Kähönen N, Jula A, et al. Tracking of serum lipid levels, blood pressure, and body mass index from childhood to adulthood: the Cardiovascular Risk in Young Finns Study. J Pediatr 2011;159:584–590. https://doi.org/10.1016/j.jpeds.2011.03.021.
3. Theodore RF, Broadbent J, Nagin D, Ambler A, Hogan S, Ramrakha S, et al. Childhood to early-midlife systolic blood pressure trajectories: early-life predictors, effect modifiers, and adult cardiovascular outcomes. Hypertension 2015;66:1108–1115. https://doi.org/10.1161/HYPERTENSIONAHA.115.05831.
4. Kim HL, Park SM, Cho IJ, Kim YM, Kim DH, Kim SH, et al. Standardized protocol of blood pressure measurement and quality control program for the Korea National Health and Nutrition Examination Survey. Clin Hypertens 2023;29:28. https://doi.org/10.1186/s40885-023-00252-7.
5. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves JW, Hill MN, et al. Recommendations for blood pressure measurement in humans: an AHA scientific statement from the Council on High Blood Pressure Research Professional and Public Education Subcommittee. J Clin Hypertens (Greenwich) 2005;7:102–109. https://doi.org/10.1111/j.1524-6175.2005.04377.x.
6. Lurbe E, Agabiti-Rosei E, Cruickshank JK, Dominiczak A, Erdine S, Hirth A, et al. 2016 European Society of Hypertension guidelines for the management of high blood pressure in children and adolescents. J Hypertens 2016;34:1887–1920. https://doi.org/10.1097/HJH.0000000000001039.
7. Flynn JT, Kaelber DC, Baker-Smith CM, Blowey D, Carroll AE, Daniels SR, et al. Clinical practice guideline for screening and management of high blood pressure in children and adolescents. Pediatrics 2017;140e20171904. https://doi.org/10.1542/peds.2017-1904.
8. Kim SH, Kim YM, Kim SH, Shin J, Lee EM. Replacing mercury sphygmomanometers with mercury-free sphygmomanometers for the National Health Survey in children: direct comparisons applying two types of mercury-free sphygmomanometer. Korean Circ J 2024;54:270–287. https://doi.org/10.4070/kcj.2023.0274.
9. Ostchega Y, Nwankwo T, Sorlie PD, Wolz M, Zipf G. Assessing the validity of the Omron HEM-907XL oscillometric blood pressure measurement device in a national survey environment. J Clin Hypertens (Greenwich) 2010;12:22–28. https://doi.org/10.1111/j.1751-7176.2009.00199.x.
10. Choi S, Kim YM, Shin J, Lim YH, Choi SY, Choi BY, et al. Comparison of the accuracy and errors of blood pressure measured by 2 types of non-mercury sphygmomanometers in an epidemiological survey. Medicine (Baltimore) 2018;97e10851. https://doi.org/10.1097/MD.0000000000010851.
11. Kim YM, Ohn DW, Kim SH, Kim DH, Park SM, Cho IJ, et al. Direct comparison of an automated oscillometric device with an electronic auscultatory device for epidemiologic survey to evaluate the prevalence of hypertension. Medicine (Baltimore) 2022;101e32299. https://doi.org/10.1097/MD.0000000000032299.
12. Kim YM, Park SH, Shin J, Sung KC, Ihm SH, Kim SH, et al. Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey. J Clin Hypertens (Greenwich) 2024;26:532–542. https://doi.org/10.1111/jch.14805.
13. Kweon S, Kim Y, Jang MJ, Kim Y, Kim K, Choi S, et al. Data resource profile: the Korea National Health and Nutrition Examination Survey (KNHANES). Int J Epidemiol 2014;43:69–77. https://doi.org/10.1093/ije/dyt228.
14. Stergiou GS, Tzamouranis D, Protogerou A, Nasothimiou E, Kapralos C. Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocol. Blood Press Monit 2008;13:299–303. https://doi.org/10.1097/MBP.0b013e3283057af6.
15. Kollias A, Ntineri A, Kyriakoulis KG, Stambolliu E, Lagou S, Boubouchairopoulou N, et al. Validation of the professional device for blood pressure measurement Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization standard. Blood Press Monit 2018;23:112–114. https://doi.org/10.1097/MBP.0000000000000307.
16. Graves JW, Tibor M, Murtagh B, Klein L, Sheps SG. The Accoson Greenlight 300, the first non-automated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults. Blood Press Monit 2004;9:13–17. https://doi.org/10.1097/00126097-200402000-00004.
17. National High Blood Pressure Education Program Working Group on High Blood Pressure in Children and Adolescents. The fourth report on the diagnosis, evaluation, and treatment of high blood pressure in children and adolescents. Pediatrics 2004;114(Suppl 2):555–576. https://doi.org/10.1542/peds.114.S2.555.
18. Kim SH, Park Y, Song YH, An HS, Shin JI, Oh JH, et al. Blood pressure reference values for normal weight korean children and adolescents: data from the Korea National Health and Nutrition Examination Survey 1998-2016: the Korean Working Group of Pediatric Hypertension. Korean Circ J 2019;49:1167–1180. https://doi.org/10.4070/kcj.2019.0075.
19. Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, et al. A universal standard for the validation of blood pressure measuring devices: association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens 2018;36:472–478. https://doi.org/10.1097/HJH.0000000000001634.
20. Araujo-Moura K, Souza LG, Mello GL, De Moraes AC. Blood pressure measurement in pediatric population: comparison between automated oscillometric devices and mercury sphygmomanometers-a systematic review and meta-analysis. Eur J Pediatr 2022;181:9–22. https://doi.org/10.1007/s00431-021-04171-3.
21. Stergiou GS, Boubouchairopoulou N, Kollias A. Accuracy of automated blood pressure measurement in children: evidence, issues, and perspectives. Hypertension 2017;69:1000–1006. https://doi.org/10.1161/HYPERTENSIONAHA.116.08553.

Article information Continued

Figure 1.

Distribution of the absolute difference between two devices. The graphs display the percentage of participants. A-diff, absolute difference; DBP, diastolic blood pressure; SBP, systolic blood pressure.

Figure 2.

The Bland-Altman plots between OD and AD for SBP and DBP according to sex. (A) SBP in the overall population, (B) DBP in the overall population, (C) SBP in boys, (D) DBP in boys, (E) SBP in girls, and (F) DBP in girls. AD, auscultatory device (Greenlight); DBP, diastolic blood pressure; OD, oscillometric device (Microlife); SBP, systolic blood pressure.

Table 1.

Characteristics of the study population according to sex

Characteristics Overall (n=251, 100%) Boys (n=121, 48.2%) Girls (n=130, 51.8%) p-value1
Age (yr) 12.6±3.3 [6–18] 12.6±3.2 [6–18] 12.5±3.5 [6–18] 0.895
Age group (yr)
 6–9 52 (20.7) 24 (19.8) 28 (21.5) 0.785
 10–12 61 (24.3) 32 (26.4) 29 (22.3)
 13–15 85 (33.9) 42 (34.7) 43 (33.1)
 16–18 53 (21.1) 23 (19.0) 30 (23.1)
AC (cm) 23.3±3.9 24.3±4.3 22.4±3.3 <0.001
Weight (kg)2 49.3±15.3 53.1±16.6 45.9±13.2 <0.001
Height (cm)2 154.2±16.0 158.2±16.2 150.6±15.0 <0.001
PR (bpm)2 79.8±11.1 78.0±11.7 81.4±10.2 0.014

Values are presented as mean±standard deviation [minimum-maximum], or number (%).

AC, arm circumference; PR, pulse rate.

1

Using the independent t-test for continuous variables and the chi-square test for categorical variables.

2

Two height and weight values and one PR value were missing.

Table 2.

Anthropometrics and blood pressure distributions by age

Variables Age (yr)
p-value1
All 6–9 10–12 13–15 16–18
AC (cm) 23.3±3.9 20.1±4.0 23.2±3.4 24.1±3.2 25.3±3.5 <0.001
Weight (kg) 49.3±15.3 31.5±10.5 47.5±11.6 55.2±11.7 60.0±12.1 <0.001
Height (cm) 154.2±16.0 130.6±9.6 151.3±10.2 163.3±7.7 166.7±8.0 <0.001
PR (bpm) 79.8±11.1 87.9±10.1 81.7±9.1 77.8±10.5 72.9±9.6 <0.001
SBP (mmHg) AD 105.4±8.7 101.7±9.0 106.3±7.6 105.6±8.7 107.5±9.1 0.004
OD 105.5±9.3 100.9±9.4 106.7±8.2 105.8±8.8 108.2±9.6 <0.001
OD–AD 0.2±5.92 –0.8±5.8 0.5±5.5 0.2±6.4 0.7±5.5 0.590
DBP (mmHg) AD 63.7±7.7 61.2±6.6 62.9±6.8 63.6±8.1 67.3±8.1 <0.001
OD 62.2±5.9 59.8±5.7 63.3±5.6 61.7±5.4 64.4±6.3 <0.001
OD–AD –1.5±6.13 –1.4±5.9 0.4±5.3 –1.9±6.9 –2.9±5.3 0.027

Values are presented as mean±standard deviation.

AC, arm circumference; PR, pulse rate; SBP, systolic blood pressure; DBP, diastolic blood pressure; AD, auscultatory device (Greenlight); OD, oscillometric device (Microlife).

1

Using one-way analysis of variance.

2

p=0.603 by paired t-test.

3

p<0.001 by paired t-test.

Table 3.

Correlations between AD and OD according to age and sex

Variables CC (95% CI)1 CCC (95% CI)
All SBP 0.79 (0.74, 0.83) 0.79 (0.74, 0.83)
DBP 0.63 (0.55, 0.70) 0.60 (0.52, 0.67)
Sex
 Boys SBP 0.77 (0.68, 0.83) 0.76 (0.67, 0.83)
DBP 0.65 (0.54, 0.75) 0.63 (0.51, 0.72)
 Girls SBP 0.77 (0.69, 0.84) 0.77 (0.69, 0.83)
DBP 0.61 (0.49, 0.71) 0.56 (0.44, 0.66)
Age (yr)
 6–9 SBP 0.81 (0.68, 0.89) 0.80 (0.68, 0.88)
DBP 0.55 (0.32, 0.72) 0.53 (0.31, 0.70)
 10–12 SBP 0.76 (0.62, 0.85) 0.75 (0.62, 0.84)
DBP 0.64 (0.46, 0.77) 0.63 (0.46, 0.76)
 13–15 SBP 0.73 (0.62, 0.82) 0.73 (0.62, 0.82)
DBP 0.54 (0.37, 0.68) 0.48 (0.32, 0.61)
 16–18 SBP 0.83 (0.71, 0.90) 0.82 (0.71, 0.90)
DBP 0.76 (0.61, 0.86) 0.68 (0.53, 0.79)
Boys (yr)
 6–9 SBP 0.80 (0.57, 0.91) 0.79 (0.56, 0.91)
DBP 0.49 (0.08, 0.76) 0.46 (0.09, 0.73)
 10–12 SBP 0.71 (0.47, 0.85) 0.71 (0.47, 0.85)
DBP 0.72 (0.48, 0.86) 0.70 (0.47, 0.84)
 13–15 SBP 0.69 (0.49, 0.83) 0.68 (0.47, 0.82)
DBP 0.56 (0.30, 0.74) 0.54 (0.29, 0.72)
 16–18 SBP 0.90 (0.76, 0.96) 0.85 (0.68, 0.93)
DBP 0.78 (0.53, 0.91) 0.73 (0.49, 0.87)
Girls (yr)
 6–9 SBP 0.77 (0.54, 0.89) 0.77 (0.54, 0.89)
DBP 0.54 (0.19, 0.77) 0.52 (0.19, 0.75)
 10–12 SBP 0.76 (0.53, 0.89) 0.75 (0.50, 0.88)
DBP 0.58 (0.25, 0.79) 0.56 (0.25, 0.77)
 13–15 SBP 0.80 (0.65, 0.89) 0.78 (0.63, 0.88)
DBP 0.58 (0.33, 0.75) 0.43 (0.22, 0.59)
 16–18 SBP 0.71 (0.46, 0.86) 0.70 (0.46, 0.85)
DBP 0.76 (0.54, 0.88) 0.61 (0.38, 0.77)

AD, auscultatory device (Greenlight); CC, Pearson's correlation coefficient; CCC, Lin's concordance correlation coefficient; CI, confidence interval; DBP, diastolic blood pressure; OD, oscillometric device (Microlife); SBP, systolic blood pressure.

1

Correlation coefficients were calculated using Pearson correlation analysis.

Table 4.

Prevalence of elevated blood pressure and hypertension according to age and sex

Variables Age (yr)
Overall
Boys
Girls
All 6–12 13–18 All 6–12 13–18 All 6–12 13–18
AD
 EBP 29 (11.6) 18 (15.9) 11 (8.0) 17 (14.1) 11 (19.6) 6 (9.2) 12 (9.2) 7 (12.3) 5 (6.9)
 HTN 12 (4.8) 8 (7.1) 4 (2.9) 5 (4.1) 4 (7.1) 1 (1.5) 7 (5.4) 4 (7.0) 3 (4.1)
OD
 EBP 25 (10.0) 19 (16.8) 6 (4.4) 16 (13.2) 11 (19.6) 5 (7.7) 9 (6.9) 8 (14.0) 1 (1.4)
 HTN 11 (4.4) 9 (8.0) 2 (1.5) 5 (4.1) 3 (5.4) 2 (3.1) 6 (4.6) 6 (10.5) 0 (0)

Values are presented as number (%).

AD, auscultatory device (Greenlight); EBP, elevated blood pressure; HTN, hypertension; OD, oscillometric device (Microlife).

Table 5.

Diagnostic agreement between AD and OD according to sex

Sex Positive concordance (n/N)1 Negative concordance (n/N)1 Agreement (95% CI, %)
PABAK (95% CI) p-value (McNemar test)
Positive Negative Overall
Overall
 EBP 16/29 213/222 55.2 (37.1, 73.3) 95.9 (93.4, 98.5) 91.2 (87.7, 94.7) 0.83 (0.76, 0.90) 0.394
 HTN 7/12 235/239 58.3 (30.4, 86.2) 98.3 (96.7, 100) 96.4 (94.1, 98.7) 0.93 (0.88, 0.97) 0.739
Boys
 EBP 10/17 98/104 58.8 (35.4, 82.2) 94.2 (89.8, 98.7) 89.3 (83.7, 94.8) 0.79 (0.68, 0.90) 0.781
 HTN 4/5 115/116 80.0 (44.9, 100) 99.1 (97.5, 100) 98.4 (96.1, 100) 0.97 (0.92, 1.00) 1.000
Girls
 EBP 6/12 115/118 50.0 (21.7, 78.3) 97.5 (94.6, 100) 93.1 (88.7, 97.4) 0.86 (0.77, 0.95) 0.317
 HTN 3/7 120/123 42.9 (6.2, 79.5) 97.6 (94.8, 100) 94.6 (90.7, 98.5) 0.89 (0.82, 0.97) 0.705

AD, auscultatory device (Greenlight); CI, confidence interval; EBP, elevated blood pressure; HTN, hypertension; OD, oscillometric device (Microlife); PABAK, prevalence-adjusted bias-adjusted kappa.

1

n/N indicates the number of participants concordantly classified by both devices divided by the total number of participants classified into the category by either device.