Device-related differences in blood pressure measurement and their impact on hypertension prevalence in Korean children and adolescents
Article information
Abstract
OBJECTIVES
Accurate blood pressure (BP) measurement is essential for reliably estimating hypertension prevalence in population-based surveys. However, evidence supporting the use of oscillometric BP devices in pediatric populations remains limited. This study compared BP measurements obtained using an oscillometric device (OD) and an auscultatory device (AD) and evaluated inter-device agreement and the impact of device choice on hypertension prevalence estimates.
METHODS
BP was measured sequentially with the AD and OD in 251 children and adolescents participating in the Korea National Health and Nutrition Examination Survey. Inter-device agreement was assessed using mean differences, limits of agreement, concordance correlation coefficients (CCCs), and regression analyses. Differences in prevalence estimates for elevated blood pressure (EBP) and hypertension were also examined.
RESULTS
Mean inter-device differences were minimal for systolic blood pressure (SBP; 0.2±5.9 mmHg) but negative for diastolic blood pressure (DBP; −1.5±6.1 mmHg), with larger discrepancies in older girls. Agreement was moderate to good for SBP (CCC=0.79) and weaker for DBP (CCC=0.60). Nevertheless, more than 85% of paired measurements were within ±10 mmHg. Prevalence estimates for EBP (11.6 vs. 10.0%) and hypertension (4.8 vs. 4.4%) were comparable between devices, with relatively high negative agreement and overall agreement.
CONCLUSIONS
Despite modest underestimation of DBP in specific subgroups, oscillometric BP measurement showed acceptable agreement with the auscultatory method and yielded similar hypertension prevalence estimates. These findings suggest that ODs may be useful for pediatric BP measurement in public health surveillance systems.
Key Message
In children and adolescents, the Microlife WatchBP Office AFIB oscillometric device showed acceptable agreement with the Accoson Greenlight 300 auscultatory device for systolic blood pressure, although diastolic blood pressure agreement was more variable in some subgroups. The two devices yielded generally comparable population-level estimates of elevated blood pressure and hypertension. These findings suggest that the Microlife device may be useful for pediatric blood pressure measurement in national health surveys.
INTRODUCTION
Accurate blood pressure (BP) measurement is fundamental to the diagnosis and surveillance of hypertension across the life course. In children and adolescents, reliable BP assessment is particularly important because elevated blood pressure (EBP) in youth is associated with increased cardiovascular risk in adulthood [1-3]. In large-scale population-based surveys, such as the Korea National Health and Nutrition Examination Survey (KNHANES), BP measurement must balance accuracy, feasibility, and standardization to ensure reliable prevalence estimates and comparability over time.
KNHANES has served as a key national health surveillance system since 1998. Historically, BP was measured using mercury sphygmomanometers, which were considered reference standards. After the Minamata Convention mandated the global phase-out of mercury-based instruments, KNHANES adopted the Accoson Greenlight 300TM as a mercury-free auscultatory device (AD) [4]. Although auscultatory methods are regarded as reference standards, they are subject to observer variability and are less practical for large-scale epidemiologic surveys [5].
Oscillometric devices (ODs) offer advantages in standardization and feasibility and are increasingly used in population-based studies. Current pediatric hypertension guidelines from the American Academy of Pediatrics (AAP) and the European Society of Hypertension (ESH) endorse validated ODs for BP screening in children and adolescents [6,7]. However, prior studies have reported discrepancies between oscillometric and auscultatory measurements, particularly for diastolic blood pressure (DBP), raising concerns about potential misclassification of BP status [8-11]. Moreover, most previous studies have focused primarily on adults or on inter-device agreement itself, whereas evidence on the impact of these differences on population-level estimates of EBP and hypertension in pediatric populations remains limited.
To address these limitations, the Microlife WatchBP Office AFIB OD has been introduced for use in KNHANES. Although adult studies have shown acceptable agreement with auscultatory methods [12], evidence regarding its performance in pediatric population-based settings remains limited. Physiological characteristics of children and adolescents, including arterial compliance and hemodynamic variability, may influence oscillometric BP measurements, particularly DBP. Additional evidence is therefore needed on inter-device agreement and the potential impact of device-related differences on BP classification in pediatric populations.
Accordingly, this study aimed to evaluate agreement between oscillometric and auscultatory BP measurements and to determine the impact of device-related differences on estimates of EBP and hypertension in a nationally representative sample of children and adolescents participating in KNHANES.
MATERIALS AND METHODS
Study participants
Between April 2023 and September 2023, 251 children and adolescents aged 6–18 years who participated in KNHANES were consecutively enrolled in this study. Each participant underwent BP measurement with both the Microlife WatchBP Office AFIB (OD) and the Accoson Greenlight 300TM (AD). Measurements were conducted at four mobile examination centers (MECs) as part of KNHANES [13]. Children were excluded if they did not complete six BP measurements with each device (three per session) or if their mid-arm circumference was <14 cm or >42 cm.
Device quality control
Calibration and quality control (QC) procedures for both devices followed the standardized KNHANES QC protocol [4]. Device accuracy was assessed using the Rigel BP-SIM NIBP Simulator® (Rigel Medical, Edison, NJ, USA) before and after field use. Daily and weekly QC procedures were also conducted at each MEC. Devices were required to meet predefined accuracy criteria, including pressure differences within ±3 mmHg for most paired readings. No device malfunction was identified during the study period.
Blood pressure measurement protocol
The OD had previously undergone validation studies in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO 81060-2:2013) [14,15], whereas the AD was validated according to the 2002 European Society of Hypertension International Protocol [16].
BP measurements were performed by four trained nurses at the MECs according to KNHANES guidelines [4]. After a 5-minute seated rest with back support and the arm positioned at heart level, participants underwent BP measurement with an appropriately sized cuff selected according to mid-arm circumference. Radial pulse was recorded for 30 seconds in participants with a regular rhythm and for 60 seconds in those with an irregular rhythm.
Measurements followed a fixed sequence to minimize recall bias: three consecutive readings were obtained with the AD, followed by three readings with the OD, with 30-second intervals between readings. For the AD, the first and fifth Korotkoff sounds were recorded as systolic blood pressure (SBP) and DBP, respectively. All measurements were obtained from the same arm under consistent conditions.
Definitions of blood pressure (BP) difference, absolute BP difference, and hypertension classification
The first reading from each device was excluded from the analysis, and the mean of the second and third readings was used. Differences in SBP and DBP (D-SBP and D-DBP) were defined as OD values minus AD values. Absolute differences (A-SBP and A-DBP) were calculated as the absolute values of these differences. EBP was defined as BP at or above the 90th percentile, and hypertension was defined as BP at or above the 95th percentile for age, sex, and height according to KNHANES reference data [17,18].
Reference for interpretation of the degree of agreement between two devices
The degree of agreement between the two devices was interpreted according to the 2018 Association for the AAMI/ESH/ISO Universal Standard for validation of BP measuring devices [19]. A device was considered to be within an acceptable range when the mean difference between OD and AD did not exceed 5±8 mmHg and at least 85% of paired measurements were within ±10 mmHg.
Statistical analysis
Categorical variables were expressed as counts and percentages, whereas continuous variables were presented as means±standard deviations. Differences between OD and AD were compared, correlations were assessed using Pearson correlation coefficients, and concordance was analyzed using Lin’s concordance correlation coefficient (CCC). Bland–Altman plots were constructed to determine limits of agreement (LOA). Using AD as the comparator, the agreement and classification performance of OD were summarized using positive agreement, negative agreement, overall agreement, and prevalence-adjusted bias-adjusted kappa (PABAK). The McNemar test was used to assess whether discordant paired classifications were symmetric. Factors associated with inter-device BP differences were examined using univariable and multivariable linear regression analyses. Subgroup analyses were performed by age group, defined as 6–9 years, 10–12 years, 13–15 years, and 16–18 years.
All statistical analyses were conducted using SAS version 9.4 (SAS Institute Inc., Cary, NC, USA), and statistical significance was set at p-value <0.05.
Ethics statement
The study was approved by the Institutional Review Board of the Korea Disease Control and Prevention Agency (approval No. 2022-11-16-R-A). Written informed consent was obtained from participants and their parents.
RESULTS
Study population
A total of 251 children and adolescents were included: 121 boys (48.2%) and 130 girls (51.8%), with a mean age of 12.6±3.3 years (Table 1). Boys had greater arm circumference (24.3±4.3 vs. 22.4±3.3 cm), body weight (53.1±16.6 vs. 45.9±13.2 kg), and height (158.2±16.2 vs. 150.6±15.0 cm) than girls (all p<0.001), whereas girls had a higher pulse rate (81.4±10.2 vs. 78.0±11.7 bpm, p=0.014).
Inter-device blood pressure differences
SBP showed minimal inter-device differences across age groups, whereas DBP measured by the OD tended to be lower than DBP measured by the AD (Table 2, Supplementary Material 1). In the overall sample, the mean SBP difference was 0.2±5.9 mmHg (p=0.603). The mean DBP difference was −1.5±6.1 mmHg (p<0.001), with larger differences observed among adolescents aged 16–18 years.
In sex-stratified analyses, SBP differences were small in both boys and girls. DBP differences were greater in girls, particularly among older adolescents (−4.0±4.4 mmHg; p<0.001).
Blood pressure (BP) agreement between BP measurement devices
Agreement between devices differed by BP component, sex, and age (Table 3). In the overall population, concordance was moderate to good for SBP (CCC, 0.79; 95% confidence interval [CI], 0.74 to 0.83) and lower for DBP (CCC, 0.60; 95% CI, 0.52 to 0.67). SBP concordance was similar in boys (CCC, 0.76) and girls (CCC, 0.77), whereas DBP concordance was 0.63 in boys and 0.56 in girls.
Across age groups, SBP concordance ranged from 0.73 to 0.82, whereas DBP concordance remained lower; the lowest value was observed in girls aged 13–15 years (CCC, 0.43). Overall, agreement was higher for SBP than for DBP across subgroups.
Absolute difference between blood pressure measurement devices
Figure 1 shows the distribution of absolute inter-device differences. More than 85% of paired measurements for both SBP and DBP were within ±10 mmHg. In the overall population, 62.6% of SBP measurements and 57.8% of DBP measurements were within ±5 mmHg.
Bland–Altman plot and limits of agreement by age and sex
Bland–Altman plots showed small mean inter-device differences in the overall population (SBP: 0.16 mmHg; DBP: −1.48 mmHg), with LOA of approximately ±12 mmHg for SBP and slightly wider ranges for DBP (Figure 2, Supplementary Material 2). Across subgroups, SBP differences remained small, whereas DBP differences were more variable, particularly in older adolescents and girls. Higher auscultatory DBP values were associated with more negative OD–AD differences.
The Bland-Altman plots between OD and AD for SBP and DBP according to sex. (A) SBP in the overall population, (B) DBP in the overall population, (C) SBP in boys, (D) DBP in boys, (E) SBP in girls, and (F) DBP in girls. AD, auscultatory device (Greenlight); DBP, diastolic blood pressure; OD, oscillometric device (Microlife); SBP, systolic blood pressure.
Factors contributing to blood pressure differences
In multivariable analysis, oscillometric SBP was independently associated with SBP differences (Supplementary Material 3). For DBP, oscillometric DBP and age were independently associated with inter-device differences. The models had limited explanatory power (SBP: R2=0.174; DBP: R2=0.077), indicating that the included variables explained only a small proportion of inter-device variability.
Implications for hypertension classification
Prevalence estimates for EBP and hypertension were similar between devices (Table 4). EBP prevalence was 11.6% with the AD and 10.0% with the OD, whereas hypertension prevalence was 4.8% and 4.4%, respectively.
Overall agreement was 91.2% for EBP (PABAK=0.83) and 96.4% for hypertension (PABAK=0.93). Similar agreement was observed in sex-specific analyses (Table 5). In the overall sample and sex-stratified analyses, OD showed high negative agreement and overall agreement with AD for both EBP and hypertension classifications. PABAK values also indicated substantial to very high agreement. However, positive agreement was relatively modest, particularly for hypertension, and confidence intervals were wide in some subgroups, suggesting uncertainty due to the small number of positive cases. McNemar tests were not statistically significant in any comparison, indicating no evidence of systematic asymmetry in discordant classifications between OD and AD.
DISCUSSION
The principal findings of this study were as follows. First, SBP measurements showed moderate to good agreement between the OD and AD across age and sex groups. Second, DBP measured by the OD tended to be lower, particularly among older adolescent girls. Third, although inter-device differences were observed, more than 85% of paired measurements were within ±10 mmHg, and population-level estimates of EBP and hypertension were similar between devices. However, positive agreement for hypertension classification was modest, indicating that agreement at the individual diagnostic level should be interpreted cautiously.
The higher agreement observed for SBP than for DBP is consistent with previous pediatric studies and reflects known technical characteristics of oscillometric measurement [20,21]. ODs estimate SBP and DBP indirectly from oscillation amplitude envelopes, and signal detection is generally more robust for SBP than for DBP, particularly in children with smaller arterial diameters and greater hemodynamic variability [20]. In this study, SBP showed stable agreement across age and sex groups, whereas DBP differences were more variable, particularly among older adolescents.
Lower concordance for DBP, especially in adolescent girls, may reflect both physiological and methodological factors. Pubertal changes, including alterations in vascular compliance and hormonal influences, as well as differences in pulse pressure amplification, may affect oscillometric signal detection during diastole [20,21]. In addition, auscultatory measurement in children is subject to variability because of difficulty identifying Korotkoff sounds and inherent measurement variability [5,20]. These factors may contribute to greater variability in DBP measurements across devices. As AD-measured DBP increased, OD–AD differences tended to become more negative; this pattern may have contributed to the lower concordance of OD-based EBP and hypertension classification among older participants.
Despite these differences, the impact on population-level hypertension classification was limited. Prevalence estimates for EBP and hypertension were generally similar between devices, although differences in agreement were observed across age and sex subgroups. Compared with the AD, the OD showed relatively high negative agreement and high overall agreement, with no statistically significant evidence of asymmetry in discordant classifications based on the McNemar test. However, positive agreement was relatively modest, particularly for hypertension classification, and CIs were wide in some subgroups. These findings suggest that oscillometric measurements may be useful for population-level surveillance, but positive classifications should be interpreted cautiously, and the two methods should not be considered fully interchangeable for individual diagnosis. This information on diagnostic concordance between the two devices may provide a useful reference for interpreting pediatric BP classifications in nationally representative health statistics.
Current pediatric hypertension guidelines from the AAP and ESH state that EBP identified by oscillometric measurement should be confirmed by auscultatory measurement before a clinical diagnosis of hypertension is established [6,7]. In this context, the present findings are more relevant to population-based surveillance and screening than to direct diagnostic interchangeability between devices. The observed inter-device differences in DBP and modest positive agreement for hypertension classification suggest that EBP identified by oscillometric measurement should be interpreted cautiously at the individual diagnostic level.
From an epidemiologic perspective, these findings are particularly relevant to population-based surveillance. National health surveys require measurement methods that ensure both accuracy and feasibility. Although auscultatory methods remain reference standards, they require trained observers and are subject to inter-observer variability [5]. In contrast, this OD offers advantages in standardization and operational efficiency while maintaining acceptable agreement with auscultatory measurements and preserving hypertension prevalence estimates.
Several limitations should be considered. First, this study was conducted in a sample of Korean children and adolescents, which may limit generalizability to other populations. Second, variability in pediatric hypertension definitions may influence prevalence estimates. Third, the relatively low prevalence of hypertension limited detailed subgroup analyses. In addition, the similar overall prevalence estimates observed in this study may partly reflect offsetting differences across age and sex subgroups. Therefore, comparability may vary depending on the age and sex composition of the target population. Fourth, BP measurements were obtained sequentially rather than using a randomized same-arm alternating measurement design, which may have introduced order effects. In particular, if oscillometric BP measurements had been obtained first, prior awareness of the displayed values could have influenced subsequent auscultatory measurements by the observer. The 2018 AAMI/ESH/ISO Universal Standard recommends a same-arm sequential alternating measurement protocol for formal validation studies. However, because this protocol was not feasible in the present study, auscultatory BP measurement was performed first to reduce the potential influence of recall bias from automated BP readings. Finally, this study should not be interpreted as a formal validation study according to the 2018 AAMI/ESH/ISO Universal Standard because the measurement protocol did not use the same-arm sequential alternating design recommended for formal device validation. Instead, the present study was designed to evaluate inter-device agreement and the potential impact of device-related differences on BP classification and hypertension prevalence estimates within the operational framework of KNHANES.
Despite these limitations, this study has important strengths. It was conducted within a nationally representative survey using standardized protocols and rigorous QC [4], which enhances the reliability and applicability of the findings for public health surveillance.
In conclusion, oscillometric BP measurements showed acceptable agreement with auscultatory measurements for SBP, whereas DBP agreement was more variable in some pediatric subgroups. Although caution is warranted when interpreting individual hypertension classifications, population-level estimates of EBP and hypertension were generally comparable between devices. These findings suggest that ODs may be useful for pediatric BP surveillance and screening in large-scale population-based studies.
Data availability
The data generated in this study are available from the corresponding author upon reasonable request.
Supplementary materials
Supplementary materials are available at https://doi.org/10.4178/epih.e2026037.
Supplementary Material 1.
Anthropometrics and blood pressure distributions by age and sex
Supplementary Material 2.
Limit of agreement (LOA) according to sex and age
Supplementary Material 3.
Factors associated with inter-device BP differences
Notes
Conflict of interest
The authors have no conflicts of interest to declare for this study.
Funding
This work was supported by the Dongguk University College of Medicine Research Fund of 2026.
Acknowledgements
None.
Author contributions
Conceptualization: Kim SH, Kim YM, Kim HL, Park SM, Cho IJ, Kim DH, Kim KI, Shin J, Oh K, Lee EM. Data curation: Kim YM, Kim SH, Lee EM. Formal analysis: Kim YM, Kim SH, Lee EM. Funding acquisition: Lee EM. Methodology: Kim YM, Kim SH, Lee EM. Project administration: Lee EM. Visualization: Kim YM, Kim SH, Lee EM. Writing – original draft: Kim YM, Kim SH, Lee EM. Writing – review & editing: Kim SH, Kim YM, Kim HL, Park SM, Cho IJ, Kim DH, Kim KI, Shin J, Oh K, Lee EM.